Lot 66Karl Storz 11101 RP2 Pharyngoscope in Case - Engineer's Report : Optical System - 1 x Broken Fiber, Angulation - Severely Crushed, Not Angled Due to Damage, Insertion Tube - No Bending Section Rubber, Light Transmission - No Fault Found, Leak Check - Leak From Insertion Tube *2816203*