Lot 65Karl Storz 11009 BCI Bronchoscope in Case - Engineer's Report : Optical System - 3 x Broken Fibers, Angulation - No Fault Found, Insertion Tube - Kinked and Leaking Due to Missing Bending Section Rubber, Light Transmission - No Fault Found, Leak Check - Leak from Insertion Tube *2192411*