Lot 94Olympus CYF-240 Video Cystoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Not Reaching Specification, To Be Adjusted, Insertion Tube - Bending Section Rubber Split, Tube Worn, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Unable to Check *2860606* (RI)