Lot 20Pentax FNL-10RBS Laryngoscope - Engineer's Report : Optical System - 1 Broken Fiber and Stain on Edge, Angulation - No Fault Found, Insertion Tube - Rubber Grip Cone Adrift, Light Transmission - No Fault Found, Leak Check - Leak at Body Area *G114418*

Click image to enlarge
Lot closed - Winning bid:N/A

March 2023 Endoscopy (#639)
10/03/2023 3:00 PM GMT - 22/03/2023 4:00 PM GMT
CLOSED
Starts Ending 22/03/2023 1:00 PM GMT

  • Pentax FNL-10RBS Laryngoscope - Engineer's Report : Optical System - 1 Broken Fiber and Stain on Edge, Angulation - No Fault Found, Insertion Tube - Rubber Grip Cone Adrift, Light Transmission - No Fault Found, Leak Check - Leak at Body Area *G114418*
  • Pentax FNL-10RBS Laryngoscope - Engineer's Report : Optical System - 1 Broken Fiber and Stain on Edge, Angulation - No Fault Found, Insertion Tube - Rubber Grip Cone Adrift, Light Transmission - No Fault Found, Leak Check - Leak at Body Area *G114418* - 2
  • Pentax FNL-10RBS Laryngoscope - Engineer's Report : Optical System - 1 Broken Fiber and Stain on Edge, Angulation - No Fault Found, Insertion Tube - Rubber Grip Cone Adrift, Light Transmission - No Fault Found, Leak Check - Leak at Body Area *G114418* - 3