Lot 62Olympus PCF-240S Video Sigmoidoscope in Case - Engineer's Report : Optical System - No Image, Angulation - Not Reaching Specification, Left / Right Brake Faulty, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2100012*

Click image to enlarge
Lot closed - Winning bid:N/A

February 2025 Endoscopy (#1083)
14/02/2025 2:00 PM GMT - 21/02/2025 3:22 PM GMT
CLOSED
Starts Ending 21/02/2025 1:00 PM GMT

  • Olympus PCF-240S Video Sigmoidoscope in Case - Engineer's Report : Optical System - No Image, Angulation - Not Reaching Specification, Left / Right Brake Faulty, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Fou
  • Olympus PCF-240S Video Sigmoidoscope in Case - Engineer's Report : Optical System - No Image, Angulation - Not Reaching Specification, Left / Right Brake Faulty, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Fou - 2
  • Olympus PCF-240S Video Sigmoidoscope in Case - Engineer's Report : Optical System - No Image, Angulation - Not Reaching Specification, Left / Right Brake Faulty, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Fou - 3
  • Olympus PCF-240S Video Sigmoidoscope in Case - Engineer's Report : Optical System - No Image, Angulation - Not Reaching Specification, Left / Right Brake Faulty, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Fou - 4