Lot 500Olympus GIF-XQ20 Gastroscope in Case - Engineer's Report : Optical System - 1 Broken Fiber, Angulation - Not Reaching Specification, Insertion Tube - Leak From Bending Section Rubber, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Leak From Bending Section Rubber *2119658* **CAMB**