Lot 32Olympus BF-P60 Bronchoscope - Engineer's Report : Optical System - Looks OK but Adapter Jammed onto Eyepiece Obscuring View, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - Minor Intermittent Leak from Channel

Click image to enlarge
Lot closed - Winning bid:N/A

June 2023 Endoscopy (#676)
31/05/2023 12:00 PM BST - 07/06/2023 3:21 PM BST
CLOSED
Starts Ending 07/06/2023 1:00 PM BST

  • Olympus BF-P60 Bronchoscope - Engineer's Report : Optical System - Looks OK but Adapter Jammed onto Eyepiece Obscuring View, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - Minor Intermittent
  • Olympus BF-P60 Bronchoscope - Engineer's Report : Optical System - Looks OK but Adapter Jammed onto Eyepiece Obscuring View, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - Minor Intermittent - 2