Lot 1726Vision Sciences TNE-5000 Rhinolaryngoscope in Case - Engineer's Report : Optical System - Untested, Angulation - Damaged Bending Section, Insertion Tube - Bending Section Damaged, No Bending Section Rubber, Light Transmission - No Fault Found, Leak Check - Untested But Suspect Leak will be Present *TE30116P* (H)