Lot 14Olympus CF-H260DL Video Colonoscope in Case - Engineer's Report : Optical System - Slightly Dark in Center, Angulation - Up / Down Loose, Requires Adjustment, Insertion Tube - No Fault Found, Light Transmission - 1 Light Bundle 60% Loss, Channels - No Fault Found, Leak Check - No Fault Found *2800262*

Click image to enlarge
Lot closed - Winning bid:N/A

May 2022 Endoscopes (#508)
18/05/2022 3:00 PM BST - 25/05/2022 3:14 PM BST
CLOSED
Starts Ending 25/05/2022 11:06 AM BST

  • Olympus CF-H260DL Video Colonoscope in Case - Engineer's Report : Optical System - Slightly Dark in Center, Angulation - Up / Down Loose, Requires Adjustment, Insertion Tube - No Fault Found, Light Transmission - 1 Light Bundle 60% Loss, Channels - No Fau
  • Olympus CF-H260DL Video Colonoscope in Case - Engineer's Report : Optical System - Slightly Dark in Center, Angulation - Up / Down Loose, Requires Adjustment, Insertion Tube - No Fault Found, Light Transmission - 1 Light Bundle 60% Loss, Channels - No Fau - 2
  • Olympus CF-H260DL Video Colonoscope in Case - Engineer's Report : Optical System - Slightly Dark in Center, Angulation - Up / Down Loose, Requires Adjustment, Insertion Tube - No Fault Found, Light Transmission - 1 Light Bundle 60% Loss, Channels - No Fau - 3
  • Olympus CF-H260DL Video Colonoscope in Case - Engineer's Report : Optical System - Slightly Dark in Center, Angulation - Up / Down Loose, Requires Adjustment, Insertion Tube - No Fault Found, Light Transmission - 1 Light Bundle 60% Loss, Channels - No Fau - 4