Lot 12Olympus BF-260 Video Bronchoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Stiff, Down Not Reaching Specification, Up Requires Adjustment, Insertion Tube - Buckled at Control Body, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Slow Leak Present *2902344*