Lot 1680Olympus CYF-2 Cystoscope in Case - Engineer's Report : Optical System - Over 20 Broken Fibers and Bad Fluid Stain Present, Angulation - Not Reaching Specification, To Be Adjusted, Insertion Tube, No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *W101059*