Catalogue Auction info

December 2021 Endoscopy (#461)
09/12/2021 2:00 PM GMT - 17/12/2021 2:08 PM GMT
CLOSED
Starts Ending 17/12/2021 1:00 PM GMT

Lot 8Olympus GIF-XQ20 Gastroscope - Engineer's Report : Optical System - 14 Broken Fibers and Stain Present, Angulation - No Fault Found, Insertion Tube - No Fault Found, Minor Bite Marks, Light Transmission - Stained, Channels - No Fault Found, Leak Check - Excessive Leak Present *2119553*

Click image to enlarge
Lot closed - Winning bid:N/A

December 2021 Endoscopy (#461)
09/12/2021 2:00 PM GMT - 17/12/2021 2:08 PM GMT
CLOSED
Starts Ending 17/12/2021 1:00 PM GMT

  • Olympus GIF-XQ20 Gastroscope - Engineer's Report : Optical System - 14 Broken Fibers and Stain Present, Angulation - No Fault Found, Insertion Tube - No Fault Found, Minor Bite Marks, Light Transmission - Stained, Channels - No Fault Found, Leak Check - E
  • Olympus GIF-XQ20 Gastroscope - Engineer's Report : Optical System - 14 Broken Fibers and Stain Present, Angulation - No Fault Found, Insertion Tube - No Fault Found, Minor Bite Marks, Light Transmission - Stained, Channels - No Fault Found, Leak Check - E - 2