Lot 23Pentax FNL-10RBS Laryngoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Crush Mark and Kink Present, Light Transmission - No Fault Found *4115976*