Lot 2Olympus CF-T10L Colonoscope - Engineer's Report : Optical System - Excessive Broken Fibers and Severe Fluid Stain, Angulation - Not Reaching Specification, Play in Controls, Bending Section Strained, Insertion Tube - Severely Kinked, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Leak Present

Click image to enlarge
Lot closed - Winning bid:N/A

December 2021 Endoscopy (#461)
09/12/2021 2:00 PM GMT - 17/12/2021 2:08 PM GMT
CLOSED
Starts Ending 17/12/2021 1:00 PM GMT

  • Olympus CF-T10L Colonoscope - Engineer's Report : Optical System - Excessive Broken Fibers and Severe Fluid Stain, Angulation - Not Reaching Specification, Play in Controls, Bending Section Strained, Insertion Tube - Severely Kinked, Light Transmission -
  • Olympus CF-T10L Colonoscope - Engineer's Report : Optical System - Excessive Broken Fibers and Severe Fluid Stain, Angulation - Not Reaching Specification, Play in Controls, Bending Section Strained, Insertion Tube - Severely Kinked, Light Transmission - - 2