Lot 15Olympus ENF-GP Pharyngoscope in Case - Engineer's Report : Optical System - 16 Broken Fibers, Angulation - Crush in Bending Section, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - Bending Section Rubber Leaking *2713288* (G)

Click image to enlarge
Lot closed - Winning bid:N/A

December 2021 Endoscopy (#461)
09/12/2021 2:00 PM GMT - 17/12/2021 2:08 PM GMT
CLOSED
Starts Ending 17/12/2021 1:00 PM GMT

  • Olympus ENF-GP Pharyngoscope in Case - Engineer's Report : Optical System - 16 Broken Fibers, Angulation - Crush in Bending Section, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - Bending Section Rubber Leaking *2713288
  • Olympus ENF-GP Pharyngoscope in Case - Engineer's Report : Optical System - 16 Broken Fibers, Angulation - Crush in Bending Section, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - Bending Section Rubber Leaking *2713288 - 2
  • Olympus ENF-GP Pharyngoscope in Case - Engineer's Report : Optical System - 16 Broken Fibers, Angulation - Crush in Bending Section, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - Bending Section Rubber Leaking *2713288 - 3