Lot 24Olympus OSF-3 Sigmoidoscope in Case - Engineer's Report : Optical System - 50+ Broken Fibers, Misty View and Condensation, Angulation - Not Reaching Specification, Insertion Tube - Badly Worn, Light Transmission - 50% Light Loss, Channels - No Fault Found, Leak Check - No Fault Found *2103674*