Lot 23Pentax FB-18BS Bronchoscope with Mains Mini Light Source (Tested Working) and Leak Tested in Case - Engineer's Report : Optical System - Approx. 6 Broken Fibers and Dirt Particles, Angulation - Strained, Not Reaching Specification, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *A01127*