Lot 94Fujinon EC-410HL Video Colonoscope in Case - Engineer's Report : Optical System - No Image, Angulation - Not Reaching Specification, To Be Adjusted, Insertion Tube - Strained, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *164D040*