Lot 8Olympus ENF Type L3 in Case - Engineer's Report : Optical System - No Broken Fibres, Angulation - OK, Insertion Tube - Torn B.S.R., Light Transmission - OK, Channels -OK , Leak Check - Leak from Tube *S/N 1410181*