Lot 50Olympus CYF-4 Cystoscope in Case - Engineer's Report : Optical System - 5 Broken Fibers and Misty View, Angulation - No Fault Found, Insertion Tube - Kinked, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Leaking (S) *W102101*