November 2023 Endoscopy 03/11/2023 1:30 PM GMT - 10/11/2023 2:14 PM GMT View Search ▼ Items per page 102550100 Search Sort by DefaultTime LeftLot#Lot NameNewestHighest PriceLowest PriceBidsViews Filter By Categories Category match ANY of the aboveALL of the above Featured Exclude closed lots Price Lot Number Results: Viewing items 1-100 of 132. Previous|12|Next Image Lot # Item Title Time Left Your Bids 1 Olympus ENF-P4 Pharyngoscope - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 800547* 1 Winning BidN/AStatusSold View details 2 Pentax FNL-10RBS Pharyngoscope Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - Kinked , Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN G11775* 1 Winning BidN/AStatusSold View details 3 Olympus ENF-P4 Pharyngoscope - Engineers Report : Optical System 0 Broken Fibers Dirt in Optics , Angulation - No Fault Found, Insertion Tube - Kinked , Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 800574* 1 Winning BidN/AStatusSold View details 4 Olympus ENF-P4 Pharyngoscope - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found, *SN 2825028* 1 Winning BidN/AStatusSold View details 5 Olympus ENF-P4 Pharyngoscope - Engineers Report : Optical System Over 40 Broken Fibers , Angulation - No Fault Found, Insertion Tube - Kinked , Light Transmission - No Fault Found, Channels - No Fault Found, *SN 1 StatusUnsold View details 6 Olympus ENF-GP Pharyngoscope - Engineers Report : Optical System 10 Broken Fibers , Angulation - No Fault Found, Insertion Tube - Kinked , Light Transmission - No Fault Found, Channels - No Fault Found, *SN 1 Winning BidN/AStatusSold View details 7 Olympus ENF-GP Pharyngoscope - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - Kinked , Light Transmission - No Fault Found, Channels - No Fault Found, *SN 700595* 1 Winning BidN/AStatusSold View details 8 Olympus ENF-P4 Pharyngoscope - Engineers Report : Optical System 3 Broken Fibers , Angulation - No Fault Found, Insertion Tube - Kinked , Light Transmission - No Fault Found, Channels - No Fault Found, *SN 800548* 1 Winning BidN/AStatusSold View details 9 Olympus ENF-P4 Pharyngoscope - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - Kinked , Light Transmission - No Fault Found, Channels - No Fault Found, *SN 800531* 1 Winning BidN/AStatusSold View details 10 Olympus ENF-P4 Pharyngoscope - Engineers Report : Optical System 3 Broken Fibers , Angulation - No Fault Found, Insertion Tube - Kinked , Light Transmission - No Fault Found, Channels - No Fault Found, *SN 800568* 1 Winning BidN/AStatusSold View details 11 Olympus ENF-P4 Pharyngoscope - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - Kinked , Light Transmission - No Fault Found, Channels - No Fault Found, *SN 1 Winning BidN/AStatusSold View details 12 Olympus ENF-P4 Pharyngoscope - Engineers Report : Optical System 1 Broken Fiber , Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, *SN 1 Winning BidN/AStatusSold View details 13 Olympus CF-Q260DL Video Colonoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2311792* 1 Winning BidN/AStatusSold View details 14 Olympus CF-Q260DL Video Colonoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Wrinkled, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 1 Winning BidN/AStatusSold View details 15 Olympus CF-Q260DL Video Colonoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2311809* 1 Winning BidN/AStatusSold View details 16 Olympus PCF-Q260AL Video Colonoscope - Engineer's Report : Optical System - No Fault Found, Angulation - Not reaching Specification - To Be Adjusted, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 23170703* 1 Winning BidN/AStatusSold View details 17 Job Lot of Scopes Including 1 x Olympus GIF-XQ Gastroscope, 1 x Olympus CF-IBW Colonoscope and 1 x Olympus BF-582 Bronchoscope (All for Spares and Repairs) 1 Winning BidN/AStatusSold View details 18 Olympus CF-240AL Video Colonoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2210499* 1 Winning BidN/AStatusSold View details 19 Olympus GIF-XQ240 Video Gastroscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2236029* 1 Winning BidN/AStatusSold View details 20 Olympus ENF-GP Pharyngoscope in Case - Engineers Report : Optical System Over 30 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN W404899* 1 Winning BidN/AStatusSold View details 21 Pentax FI-10BS Laryngoscope in Case (Dismantled Condition) - Engineers Report : Optical System No Focus , Angulation - Strained, Insertion Tube - Torn *sn a01186*(IPS) 1 Winning BidN/AStatusSold View details 22 Olympus CF-Q240L Colonoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - Bending Section Strained , Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2900347* 1 Winning BidN/AStatusSold View details 23 Olympus GIF-XQ240 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Bending Section Strained, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2033710* 1 Winning BidN/AStatusSold View details 24 Olympus GIF-XQ240 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Not Reaching Specification - To Be Adjusted, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2033649* 1 Winning BidN/AStatusSold View details 25 Olympus GIF-XQ240 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Bending Section Strained, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2032195* 1 Winning BidN/AStatusSold View details 26 Olympus GIF-XQ240 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Bending Section Strained, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2235251* 1 Winning BidN/AStatusSold View details 27 Olympus GIF-XQ240 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Bending Section Strained, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2033727* 1 Winning BidN/AStatusSold View details 28 Olympus CF-H260DL Video Colonoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2900398* 1 Winning BidN/AStatusSold View details 29 Olympus CYF-5 Cystoscope In Case - Engineer's Report : Optical System - Bad Fluid Stain , Angulation - No Fault Found, Insertion Tube - Badly Kinked, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 500924* 1 StatusUnsold View details 30 Olympus ENF-P4 Pharyngoscope - Engineers Report : Optical System 2 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found, *SN 011373* 1 Winning BidN/AStatusSold View details 31 Olympus CF-H260DL Video Colonoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2900450* 1 Winning BidN/AStatusSold View details 32 Olympus GIF-H260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2141287* 1 Winning BidN/AStatusSold View details 33 Olympus GIF-H260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Badly Kinked, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2140801* 1 Winning BidN/AStatusSold View details 34 Olympus CF-H260DL Video Colonoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2000480* 1 Winning BidN/AStatusSold View details 35 Olympus CF-H260DL Video Colonoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2100555* 1 Winning BidN/AStatusSold View details 36 Olympus CYF-240 Video Cystoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2000750* 1 Winning BidN/AStatusSold View details 37 Olympus CF-H260DL Video Colonoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - Not Reaching Specification - To Be Adjusted , Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2900424*( 1 Winning BidN/AStatusSold View details 38 Olympus GIF-2T240 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Bending Section Strained , Insertion Tube - Minor Kink , Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found 1 Winning BidN/AStatusSold View details 39 Olympus ENF-P4 Pharyngoscope - Engineers Report : Optical System Minor Stain , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found, *SN 512478* 1 Winning BidN/AStatusSold View details 40 Olympus LF GP Laryngoscope in Case - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found *SN 1000932* 1 Winning BidN/AStatusSold View details 41 Olympus GIF-H260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2243098* 1 Winning BidN/AStatusSold View details 42 Olympus BF-1T60 Bronchoscope in Case - Engineers Report : Optical System Minor Stain , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found *SN 2801422* 1 Winning BidN/AStatusSold View details 43 Olympus GIF-H260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Badly Kinked, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2142233* 1 Winning BidN/AStatusSold View details 44 Olympus CYF-240 Video Cystoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Minor Crush, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2000741* 1 Winning BidN/AStatusSold View details 45 Olympus GIF-H260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2146790* 1 Winning BidN/AStatusSold View details 46 Olympus GIF-H260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2243191* 1 Winning BidN/AStatusSold View details 47 Olympus CF-H260DL Video Colonoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2900444* 1 Winning BidN/AStatusSold View details 48 Olympus GIF-H260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2243191* 1 Winning BidN/AStatusSold View details 49 Olympus GIF-H260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2141290* 1 Winning BidN/AStatusSold View details 50 Olympus LF-TP Laryngoscope in Case - Engineers Report : Optical System Excessive Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found *SN 700531* 1 StatusUnsold View details 51 Olympus CF-H260DL Video Colonoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2700675* 1 Winning BidN/AStatusSold View details 52 Olympus ENF-GP laryngoscope in Case - Engineers Report : Optical System Fluid Stain , Angulation - No Fault Found, Insertion Tube - Badly Kinked , Light Transmission - No Fault Found, Channels - No Fault Found *SN 1 Winning BidN/AStatusSold View details 53 Olympus BF-260 Video Bronchoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Minor Crush, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2002699* 1 Winning BidN/AStatusSold View details 54 Olympus CYF-240 Video Cystoscope in Case - Engineer's Report : Optical System - No Image, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 200745* 1 StatusUnsold View details 55 Olympus LF GP Laryngoscope in Case - Engineers Report : Optical System - Fluid Stain , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found *SN 1201919* 1 Winning BidN/AStatusSold View details 56 Olympus LF-TP Laryngoscope in Case - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found *SN 1102025* 1 Winning BidN/AStatusSold View details 57 Olympus LF-TP Laryngoscope in Case - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found *SN 700524* 1 Winning BidN/AStatusSold View details 58 Olympus BF-P60 Bronchoscope in Case - Engineers Report : Optical System 2 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found *SN 2802146* 1 Winning BidN/AStatusSold View details 59 Olympus CF-H260DL Video Colonoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2900442* 1 Winning BidN/AStatusSold View details 60 Olympus TJF-260V Video Duodenoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2101502 1 Winning BidN/AStatusSold View details 61 Olympus LF-TP Laryngoscope in Case - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found *SN 700522* 1 Winning BidN/AStatusSold View details 62 Olympus BF-260 Video Bronchoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2102984* 1 Winning BidN/AStatusSold View details 63 Olympus CYF-240 Video Cystoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2000749* 1 Winning BidN/AStatusSold View details 64 Olympus GIF-H260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2243096* 1 Winning BidN/AStatusSold View details 65 Olympus LF-GP Laryngoscope in Case - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found *SN 50064* 1 Winning BidN/AStatusSold View details 66 Olympus CYF-240 Video Cystoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Minor Crush, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2000747* 1 Winning BidN/AStatusSold View details 67 Olympus CYF-240 Video Cystoscope in Case - Engineer's Report : Optical System - Intermittent Image, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2000743* 1 StatusUnsold View details 68 Olympus LF-TP Laryngoscope in Case - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found *SN 7000530* 1 Winning BidN/AStatusSold View details 69 Olympus ENF-P4 Pharyngoscope - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - Minor Kink , Light Transmission - No Fault Found, Channels - No Fault Found *SN 800524* 1 Winning BidN/AStatusSold View details 70 Olympus CF-Q260DL Video Colonoscope in Case - Engineer's Report : Optical System - Freeze Frame Inoperable , Angulation - Not Reaching Specification- To Be Adjusted, Insertion Tube - Wrinkled, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2510253* 1 Winning BidN/AStatusSold View details 71 Olympus CYF-240 Video Cystoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2000748* 1 Winning BidN/AStatusSold View details 72 Olympus BF-P60 Bronchoscope in Case - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found *SN 2802070* 1 Winning BidN/AStatusSold View details 73 Olympus GIF-H260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2040495* 1 Winning BidN/AStatusSold View details 74 Olympus CF-H260DL Video Colonoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2900449* 1 Winning BidN/AStatusSold View details 75 Olympus CF-H260DL Video Colonoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2900451* 1 Winning BidN/AStatusSold View details 76 Olympus CF-H260DL Video Colonoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2900446* 1 Winning BidN/AStatusSold View details 77 Olympus BF-1T260 Video Bronchoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2001709* 1 Winning BidN/AStatusSold View details 78 Olympus CF-H260DL Video Colonoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2200676* 1 Winning BidN/AStatusSold View details 79 Olympus CF-H260DL Video Colonoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2900445* 1 Winning BidN/AStatusSold View details 80 Olympus BF-260 Video Bronchoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2102297* 1 Winning BidN/AStatusSold View details 81 Olympus GIF-H260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - Tube Has Minor Kink, Channels - No Fault Found, Leak Check - No Fault Found *SN 2141280* 1 Winning BidN/AStatusSold View details 82 Olympus CF-H260DL Video Colonoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2260677* 1 Winning BidN/AStatusSold View details 83 Olympus CYF-240 Video Cystoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 200754* 1 Winning BidN/AStatusSold View details 84 Olympus CF-H260DL Video Colonoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2200678* 1 Winning BidN/AStatusSold View details 85 Olympus GIF-H260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2142373* 1 Winning BidN/AStatusSold View details 86 Karl Storz AIDA NEO Control Unit *HDD Removed* (Powers Up) *SN KSK21367* 1 StatusUnsold View details 87 Olympus PCF-240L Video Colonoscope in Case - Engineer's Report : Optical System - Misty View , Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN2900101* 1 Winning BidN/AStatusSold View details 88 Olympus MH-908 Ultrasonic Probe In Case - Engineer's Report : Optical System - Unable to Check , Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN1500036* 1 StatusUnsold View details 89 Pentax FNL-10RBS Laryngoscope In Case - Engineers Report : Optical System 2 Broken Fibers, Angulation - Down Angulation Broken, Insertion Tube - Kinked, Light Transmission - No Fault Found, Leak Check - No Fault Found *SN G111558* 1 StatusUnsold View details 90 Olympus CYF-4 Cystoscope in Case - Engineers Report : Optical System Misty View Over 50 Broken Fibers , Angulation - No Fault Found, Insertion Tube - Crush Marks, Stained, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 203740* 1 StatusUnsold View details 91 Olympus HYF Type XP Hysteroscope In Case - Engineers Report : Optical System Excess Broken Fibers , Angulation - No Fault Found, Insertion Tube - Dull, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found 1 Winning BidN/AStatusSold View details 92 Olympus GF-UM2000 Ultrasound Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Unable to Check Ultrasound Image, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 1310304* 1 StatusUnsold View details 93 Olympus CYF-240 Video Cystoscope *Mfd 2014* in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Not Reaching Specification - No Fault Found , Insertion Tube - Bending Section Rubber Split, Tube Worn, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Unable to Check *SN 2860606* 1 StatusUnsold View details 94 Olympus CYF-240 Video Cystoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Not Reaching Specification, To Be Adjusted, Insertion Tube - Bending Section Rubber Split, Tube Worn, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Unable to Check *2860606* (RI) 1 StatusUnsold View details 95 Smith & Nephew Ref 7205928 30 Degree Autoclavable Video Arthroscope - Engineer's Report : Optical System - Unable to Test, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN XH30187* 1 Winning BidN/AStatusSold View details 96 Smith & Nephew Ref 7210716 30 Degree Multimode Video Arthroscope - Engineer's Report : Optical System - Unable to Test, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN XR3960* 1 Winning BidN/AStatusSold View details 97 Pentax FNL-10RBS Pharyngoscope - Engineer's Report : Optical System - 4 Broken Fibres, Angulation - Kinked and Perforated, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - NLeaking Insertion Tube *SN A010291* 1 Winning BidN/AStatusSold View details 98 Pentax FNL-10RBS Rhinolaryngoscope - Engineer's Report : Optical System - Fluid and Dirt Stain , Angulation - No Fault Found, Insertion Tube - Badly Kinked, Light Transmission - No Fault Found, Leak Check - Tube Leak *SN G111845* 1 StatusUnsold View details 99 Xion EF-N Pharyngoscope - Engineer's Report : Optical System - Fluid Stain, Angulation - Stained, Insertion Tube - Kinks, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 05725* 1 Winning BidN/AStatusSold View details 100 Olympus TJF-260V Video Duodenoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2101360* 1 Winning BidN/AStatusSold View details Items per page 102550100 Previous|12|Next Previous 12 Next