BRITISH MEDICAL AUCTIONS October 2023 Endoscopy Equipment (#767) 09/10/2023 3:00 PM BST - 16/10/2023 2:12 PM BST CLOSED Starts Ending 16/10/2023 10:04 AM BST Auction info BRITISH MEDICAL AUCTIONS View Search ▼ Items per page 102550100 Search Sort by DefaultTime LeftLot#Lot NameNewestHighest PriceLowest PriceBidsViews Filter By Categories Category match ANY of the aboveALL of the above Featured Exclude closed lots Price Lot Number Results: Viewing items 1-100 of 127. Previous|12|Next Lot 1 Olympus ENF Type GP Rhinolaryngoscope In Case - Engineers Report : Optical System - 0 Broken Fibres, Angulation - No Fault Found, Insertion Tube - Minor Kink, Light Transmission - No Fault Found, Leak Check - No Fault Found *SN 700719* View details Winning BidN/A Lot 2 Olympus ENF-GP laryngoscope in Case - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - Minor Kink, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found View details Winning BidN/A Lot 3 Olympus ENF-T3 Pharyngoscope in Case - Engineers Report : Optical System 2 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 900198* View details Winning BidN/A Lot 4 Olympus ENF-T3 Pharyngoscope in Case - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 900197* View details Winning BidN/A Lot 5 Olympus ENF-T3 Pharyngoscope in Case - Engineers Report : Optical System 1 Broken Fibre, Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Leak Check - No Fault Found, Channels - No Fault Found *SN 900196* View details Winning BidN/A Lot 6 Olympus GIF-1T240 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2900110* View details Winning BidN/A Lot 7 Olympus GIF-2T240 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Not Reaching Specification - To Be Adjusted, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2630878* View details Winning BidN/A Lot 8 Olympus TJF-260V Video Duodenoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2700349* View details Winning BidN/A Lot 9 Olympus PCF-240S Video Sigmoidoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2600238* View details Winning BidN/A Lot 10 Olympus TJF-260V Video Duodenoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2700338* View details Winning BidN/A Lot 11 Olympus JF-260V Video Duodenoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found View details Winning BidN/A Lot 12 Olympus ENF-GP laryngoscope in Case - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - Minor Kink, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 700736* View details Winning BidN/A Lot 13 Karl Storz Endoskope 203020 20 SCB equimat Unit (Powers Up with E001 Message) View details Winning BidN/A Lot 14 Olympus ENF-GP laryngoscope in Case - Engineers Report : Optical System 3 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found View details Winning BidN/A Lot 15 Olympus CYF-4 Cystoscope in Case - Engineers Report : Optical System Misty View Over 50 Broken Fibers , Angulation - No Fault Found, Insertion Tube - Crush Marks, Stained, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 203740* View details StatusUnsold Lot 16 Olympus HYF Type XP Hysteroscope In Case - Engineers Report : Optical System Excess Broken Fibers , Angulation - No Fault Found, Insertion Tube - Dull, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found View details StatusUnsold Lot 17 Olympus GIF-Q240Z Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN2200185* View details Winning BidN/A Lot 18 Olympus CF-H260DL Video Colonoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2300782* View details Winning BidN/A Lot 19 Olympus TJF-240 Video Duodenoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2300834* View details Winning BidN/A Lot 20 Olympus LF GP Laryngoscope in Case - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 1312895* View details Winning BidN/A Lot 21 Olympus GF-UC240P Ultrasound Video Gastroscope in Case - Engineer's Report : Optical System - Unable to Check, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found View details Winning BidN/A Lot 22 Olympus MH-908 Ultrasonic Probe In Case - Engineer's Report : Optical System - Unable to Check , Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN1500036* View details StatusUnsold Lot 23 Olympus GF-UM2000 Ultrasound Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Unable to Check Ultrasound Image, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 1310304* View details StatusUnsold Lot 24 Olympus TJF-240 Video Duodenoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Stained, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found View details Winning BidN/A Lot 25 Olympus TJF-240 Video Duodenoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Crush Mark, Stained, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2030248* View details Winning BidN/A Lot 26 Olympus PCF-240L Video Colonoscope in Case - Engineer's Report : Optical System - Misty View , Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN2900101* View details StatusUnsold Lot 27 Olympus TJF 200 Duodenoscope In Case - Engineer's Report : Optical System - Not Reaching Specification / To Be Adjusted, Angulation - No Fault Found, Insertion Tube - Stained, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2600365* View details Winning BidN/A Lot 28 Karl Storz 11302 BDD2 Video Laryngoscope in Case - Engineers Report : Optical System Unable to Check, Angulation - No Fault Found, Insertion Tube - Badly Crushed , Light Transmission - 25% Loss, Channels - No Fault Found, Leak Check - No Fault Found View details Winning BidN/A Lot 29 Olympus LF GP Laryngoscope in Case - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 700329* View details Winning BidN/A Lot 30 Olympus LF GP Laryngoscope in Case - Engineers Report : Optical System 0 Broken Fibers - Fluid Stain , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Intermittent Body Leak *SN 1314687* View details Winning BidN/A Lot 31 Pentax ECY-1575K Cystoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN G120049* View details Winning BidN/A Lot 32 Karl Storz Endoskope 207121 20 Multidrive Morcellator / Electrosurgical Console (Powers Up) View details Winning BidN/A Lot 33 XGIF-Q240FZ Prototype Gastroscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found View details Winning BidN/A Lot 34 Olympus CF-240DL Video Colonoscope in Case (Case Missing Handle) - Engineer's Report : Optical System - No Fault Found, Angulation - Broken Up Angulation, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2300146* View details Winning BidN/A Lot 35 Olympus BF-240 Video Bronchoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Kinked, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2800313 View details Winning BidN/A Lot 36 Olympus BF-260 Video Bronchoscope In Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2103111* View details Winning BidN/A Lot 37 Olympus BF-240 Video Bronchoscope In Case - Engineer's Report : Optical System - Misty Image, Angulation - No Fault Found, Insertion Tube - Kinked, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 1340963* View details Winning BidN/A Lot 38 Karl Storz 11101RP2 Pharyngoscope in Case - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found View details Winning BidN/A Lot 39 Karl Storz 11101RP2 Pharyngoscope in Case - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found View details Winning BidN/A Lot 40 Karl Storz AIDA NEO Control Unit *HDD Removed* (Powers Up) *SN KSK21367* View details StatusUnsold Lot 41 Olympus ENF-P4 Pharyngoscope In Case - Engineers Report : Optical System 3 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 1126639* View details Winning BidN/A Lot 42 Olympus ENF-P4 Pharyngoscope In Case - Engineers Report : Optical System 1 Broken Fibers , Angulation - No Fault Found, Insertion Tube - Minor Kink, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 305988* View details Winning BidN/A Lot 43 Olympus ENF-P4 Pharyngoscope In Case - Engineers Report : Optical System 0 Broken Fibers Fluid Stain , Angulation - No Fault Found, Insertion Tube - Minor Kink, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Leaking Body *SN 700260* View details Winning BidN/A Lot 44 Olympus ENF-P4 Pharyngoscope In Case - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN View details Winning BidN/A Lot 45 Olympus ENF-P4 Pharyngoscope In Case - Engineers Report : Optical System 0 Broken Fibers , Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 125607* View details Winning BidN/A Lot 46 Karl Storz Endoskope 207121 20 Multidrive Morcellator / Electrosurgical Console (Powers Up) View details Winning BidN/A Lot 47 Smith & Nephew Ref 7210716 30 Degree Multimode Video Arthroscope - Engineer's Report : Optical System - Unable to Test, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found View details Winning BidN/A Lot 48 Smith & Nephew Ref 7210716 30 Degree Multimode Video Arthroscope - Engineer's Report : Optical System - Unable to Test, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN XR3482* View details Winning BidN/A Lot 49 Olympus GIF 200 Scope - Engineer's Report : Optical System - No Fault Found, Angulation - Strained, Insertion Tube - Badly Worn and Stained, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2000027* View details Winning BidN/A Lot 50 Olympus JF-10 Duodenoscope - Engineer's Report : Optical System - Excess Broken Fibres, Angulation - Strained, Insertion Tube - Severely Kinked and Torn, Light Transmission - Poor, Channels - No Fault Found, Leak Check -Leaking Tube *SN 1300107* View details Winning BidN/A Lot 51 Olympus GIF Type P20 Gastroscope - Engineer's Report : Optical System - Excess Broken Fibres, Stained Image, Angulation - Strained, Insertion Tube - Collapsed, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Tube Leaking *SN 2933582* View details Winning BidN/A Lot 52 Olympus CF-10L Colonoscope - Engineer's Report : Optical System - 20 Broken Fibres, Fluid Stain , Angulation - Strained, Insertion Tube - Badle Strained, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found View details Winning BidN/A Lot 53 Vision Sciences CST-5000 Video Cystoscope - Engineers Report : Optical System Unable to Check, Angulation - No Fault Found, Insertion Tube - Severed at Body , Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found View details Winning BidN/A Lot 54 Vision Sciences TNE-2000 Rhinolaryngoscope - Engineers Report : Optical System Total Image Losss , Insertion Tube - Severed, Light Transmission - Tube Severed View details Winning BidN/A Lot 55 Olympus CF-Q260DL Video Colonoscope - Engineer's Report : Optical System - No Fault Found, Angulation - Strained, Insertion Tube - Badly Kinked, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2211526* View details Winning BidN/A Lot 56 Olympus CF-Q260DL Video Colonoscope - Engineer's Report : Optical System - No Fault Found, Angulation - Strained, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found View details Winning BidN/A Lot 57 Olympus CF-Q240L Colonoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Rubber Grip Split, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2800155* View details Winning BidN/A Lot 58 Olympus CF-Q260DL Video Colonoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2211527* View details Winning BidN/A Lot 59 Olympus GIF-Q260 Video Gastroscope - Engineer's Report : Optical System - Intermittent Image, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2532986* View details Winning BidN/A Lot 60 Olympus CF-Q260DL Video Colonoscope - Engineer's Report : Optical System - No Fault Found, Angulation - Badly Stained, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2811097* View details Winning BidN/A Lot 61 Olympus GIF-Q260 Video Gastroscope - Engineer's Report : Optical System - No Fault Found, Angulation - Not Reaching Specification To Be Adjusted, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2857660* View details Winning BidN/A Lot 62 Olympus BF-240 Video Bronchoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Badle Kinked, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 1940254* View details Winning BidN/A Lot 63 Olympus GIF-Q260 Video Gastroscope - Engineer's Report : Optical System - No Fault Found, Angulation - Badle Strained, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2857676* View details Winning BidN/A Lot 64 Olympus GIF-Q260 Video Gastroscope - Engineer's Report : Optical System - No Fault Found, Angulation - Not Reaching Specification To Be Adjusted, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found View details Winning BidN/A Lot 65 Olympus CF-Q260DL Video Colonoscope - Engineer's Report : Optical System - No Fault Found, Angulation - L /R Movement Strained, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2810959* View details Winning BidN/A Lot 66 Olympus CF-Q260DL Video Colonoscope - Engineer's Report : Optical System - No Fault Found, Angulation - Strained, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2211523* View details Winning BidN/A Lot 67 Olympus CF-Q260DL Video Colonoscope - Engineer's Report : Optical System - No Fault Found, Angulation - Strained, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2810954* View details Winning BidN/A Lot 68 Olympus ENF-T3 Pharyngoscope - Engineers Report : Optical System 0 Broken Fibre, Angulation - No Fault Found, Insertion Tube - No Fault Found , Light Transmission - No Fault Found, Leak Check - No Fault Found, Channels - No Fault Found *SN 900199* View details Winning BidN/A Lot 69 Smith & Nephew Ref 7210716 30 Degree Multimode Video Arthroscope - Engineer's Report : Optical System - Unable to Test, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN XR3960* View details StatusUnsold Lot 70 Smith & Nephew Ref 7205928 30 Degree Autoclavable Video Arthroscope - Engineer's Report : Optical System - Unable to Test, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN XH30187* View details StatusUnsold Lot 71 Karl Storz 11302 BDD2 Flexible Video Laryngoscope in Carry Case with 1 x Karl Storz Endoskope Lead Tester - Engineer's Report : Optical System - Excessive Broken Fibres, Angulation - No Fault Found, Insertion Tube - Crush Marks View details Winning BidN/A Lot 72 Pentax FNL-10RBS Laryngoscope In Case - Engineers Report : Optical System 2 Broken Fibers, Angulation - Down Angulation Broken, Insertion Tube - Kinked, Light Transmission - No Fault Found, Leak Check - No Fault Found *SN G111558* View details StatusUnsold Lot 73 Olympus CHF-CB30L Choledocoscope in Case - Engineers Report : Optical System Total Image Loss, Angulation - Strained, Insertion Tube - Indentations, Light Transmission - No Fault Found, Leak Check - No Fault Found *SN W800917* View details Winning BidN/A Lot 74 Pentax EC-3885M Colonoscope Engineer's Report : Optical System - Unable to Check, Angulation - Strained, Insertion Tube - Light Scuffing, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN A11014* View details Winning BidN/A Lot 75 Olympus CYF-240 Video Cystoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 2800560* View details Winning BidN/A Lot 76 Olympus CYF-240 Video Cystoscope *Mfd 2014* in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Not Reaching Specification - No Fault Found , Insertion Tube - Bending Section Rubber Split, Tube Worn, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Unable to Check *SN 2860606* View details StatusUnsold Lot 77 Olympus CYF-240 Video Cystoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Not Reaching Specification, To Be Adjusted, Insertion Tube - Bending Section Rubber Split, Tube Worn, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Unable to Check *2860606* (RI) View details StatusUnsold Lot 78 Pentax FNL-10RBS Rhinolaryngoscope - Engineer's Report : Optical System - Fluid and Dirt Stain , Angulation - No Fault Found, Insertion Tube - Badly Kinked, Light Transmission - No Fault Found, Leak Check - Tube Leak *SN G111845* View details StatusUnsold Lot 79 Xion EF-N Pharyngoscope - Engineer's Report : Optical System - Fluid Stain, Angulation - Stained, Insertion Tube - Kinks, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *SN 05725* View details StatusUnsold Lot 80 Pentax FNL-10RBS Pharyngoscope - Engineer's Report : Optical System - 4 Broken Fibres, Angulation - Kinked and Perforated, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - NLeaking Insertion Tube *SN A010291* View details StatusUnsold Lot 81 Olympus CYF-3 Cystoscope in Case - Engineer's Report : Optical System - Excessive Broken Fibres, Angulation - Stained, Insertion Tube - Ripped Sheath, Light Transmission - Excessive Broken Fibres *SN 905331* View details Winning BidN/A Lot 82 Smith & Nephew 0L-1 Endoscope - Engineer's Report : Optical System - Excessive Broken Fibres, Angulation - No Fault Found, Insertion Tube - Cone Broken, Light Transmission - No Fault Found View details Winning BidN/A Lot 83 Olympus GF-UC240 Ultrasound Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found (Ultrasound Untested) *SN 1000012* View details Winning BidN/A Lot 84 Olympus CLV-S190 Visera Elite Light Source (Powers Up) View details StatusUnsold Lot 85 Karl Storz Endoskope Model 470D Cold Light Fountain Light Source (Powers Up) *SN BN0000020014* View details StatusUnsold Lot 86 Olympus CLH-SC Halogen Light Source (Powers Up) *SN 7525905* View details Winning BidN/A Lot 87 DP Medical Stroboview 2000 Camera Control Unit View details StatusUnsold Lot 88 Olympus UCR C02 Insufflator Unit (Powers Up) *SN 7226782* View details Winning BidN/A Lot 89 Olympus HYF Type XP Hysteroscope *Mfd 2010* In Case - Engineer's Report : Optical System - Excessive Dirt and Stained, Light Transmission - No Fault Found, Angulation - Strained, Insertion Tube - Bent Rubber Section, Leak Check- No Fault Found *SN W102120* View details Winning BidN/A Lot 90 Non Fibre Optic Nasopharyngoscope View details Winning BidN/A Lot 91 Karl Storz H3 Image1 HD Camera Ref 22220150 *Untested* View details Winning BidN/A Lot 92 Karl Storz H3-Z Image1 HD Camera Ref 2222055 *Untested* View details Winning BidN/A Lot 93 Karl Storz H3-ZA Image 1 HD Camera Ref 22220061 *Untested* View details Winning BidN/A Lot 94 Karl Storz Endoskope 202121 20 telecam SL ntsc Camera Control Console (No Power) View details Winning BidN/A Lot 95 Stryker 40L Highflow Insufflator (Powers Up) View details Winning BidN/A Lot 96 Stryker L9000 LED Light Source (No Power) View details Winning BidN/A Lot 97 Olympus OEV261H High Definition LCD Endoscopy Monitor with Power Supply (Powers Up, Light Scratches On Screen - See Photo) View details Winning BidN/A Lot 98 5 x Light Cables Ref 07-3060 View details Winning BidN/A Lot 99 Stryker 1288-210-105 1288HD Camera View details Winning BidN/A Lot 100 Olympus Evis Lucera MAJ-1154 Pigtail Connector View details Winning BidN/A Items per page 102550100 Previous|12|Next Previous 12 Next Previous 12 Next