July 2022 Two Day Live Medical Equipment Auction July 7, 2022 View Search ▼ Items per page 102550100 Search Sort by DefaultTime LeftLot#Lot NameNewestHighest PriceLowest PriceBidsViews Filter By Categories Category match ANY of the aboveALL of the above Featured Exclude closed lots Price Lot Number Results: Viewing items 101-110 of 110. Previous|12|Next Image Lot # Item Title Time Left Your Bids 1719 Olympus LF-GP Laryngoscope in Case - Engineer's Report : Optical System - 7 Broken Fibers and Fluid / Dirt Stain Present, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Slow Tube Leak Present *1413685* 1 Winning BidN/AStatusSold View details 1720 Baxter 1 Meter Ultrafine Scope in Case - Engineer's Report : Optical System - Blurred, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - Unable to Check 1 StatusUnsold View details 1721 Olympus CHF-PQ20 Choledocoscope - Engineer's Report : Optical System - Blurred Image, Angulation - No Fault Found, Insertion Tube - Kinked, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Leaking Tube *2420446* 1 StatusUnsold View details 1722 Olympus CYF-240 Video Cystoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *5707-894* 1 Winning BidN/AStatusSold View details 1723 Olympus CYF-240 Video Cystoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Crush and Kink Marks Present, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *5707-894* 1 Winning BidN/AStatusSold View details 1724 Olympus LS-10R Teaching Attachment in Case 1 Winning BidN/AStatusSold View details 1725 Vision Sciences TNE-2000 Rhinolaryngoscope in Case - Engineer's Report : Optical System - Multiple Broken Fibers, Angulation - Snaking, Insertion Tube - Multiple Crushed Sections, Light Transmission - Loss of Light, Channels - Blocked, Leak Check - No Fault Found, Other Comments - Light Guide Coating Damaged *T02745* (H) 1 StatusUnsold View details 1726 Vision Sciences TNE-5000 Rhinolaryngoscope in Case - Engineer's Report : Optical System - Untested, Angulation - Damaged Bending Section, Insertion Tube - Bending Section Damaged, No Bending Section Rubber, Light Transmission - No Fault Found, Leak Check - Untested But Suspect Leak will be Present *TE30116P* (H) 1 Winning BidN/AStatusSold View details 1727 Vision Sciences TNE-2000 Rhinolaryngoscope in Case - Engineer's Report : Optical System - 2 Broken Fibers, Angulation - No Fault Found, Insertion Tube - Badly Crushed, Light Transmission - No Fault Found, Leak Check - Light Guide Tube Torn and Leaking *T0226P* (H) 1 Winning BidN/AStatusSold View details 1728 Pentax FI-13P Laryngoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *A01002* 1 StatusUnsold View details Items per page 102550100 Previous|12|Next Previous 12 Next