BRITISH MEDICAL AUCTIONS May 2022 Endoscopes (#508) 18/05/2022 3:00 PM BST - 25/05/2022 3:14 PM BST CLOSED Starts Ending 25/05/2022 11:06 AM BST Auction info BRITISH MEDICAL AUCTIONS View Search ▼ Items per page 102550100 Search Sort by DefaultTime LeftLot#Lot NameNewestHighest PriceLowest PriceBidsViews Filter By Categories Category match ANY of the aboveALL of the above Featured Exclude closed lots Price Lot Number Results: Viewing items 1-100 of 141. Previous|12|Next Lot 1 Olympus LF-GP Laryngoscope with Olympus MAJ-524 Portable Light Source - Engineer's Report : Optical System - No Fault Found, Angulation - Out of Spec, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *1101271* View details Winning BidN/A Lot 2 Olympus CF-H260DL Video Colonoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2300799* View details Winning BidN/A Lot 3 Olympus LF-TP Laryngoscope with Olympus MAJ-524 Portable Light Source (Powers Up) in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *W800556* View details Winning BidN/A Lot 4 Olympus ENF Type P4 Rhinolaryngoscope in Case - Engineer's Report : Optical System - Poor, Angulation - Damaged, Insertion Tube - Crushed at Control Body, Light Transmission - 100% Loss, Channels - Crushed, Leak Check - Failed, Other Comments - Scratched Control Body *W011412* View details Winning BidN/A Lot 5 Olympus GIF-N260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Leak Present at Bending Section Rubber *2610585* View details StatusUnsold Lot 6 Karl Storz 13801NKS Video Gastroscope in Case - Engineer's Report : Optical System - Untested, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault FOund, Leak Check - Untested View details StatusUnsold Lot 7 VES Xion BXEFV14 Bronchoscope in Case - Engineer's Report : Optical System - Multiple Broken Fibers, Angulation - Damaged, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Leak Present *16300* View details StatusUnsold Lot 8 Pentax EC-3870LK Video Colonoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Untested, Other Comments - Small Crush to Umbilical about Half Way *A120252* View details Winning BidN/A Lot 9 Olympus GIF-N260 Video Gastroscope with EndoMed Leak Tester in Case - Engineer's Report : Optical System - Cloudy Image, Angulation - Down Requires Adjustment, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2610551* View details StatusUnsold Lot 10 Olympus CF-P10L Choledocoscope in Case - Engineer's Report : Optical System - Cloudy, Angulation - Requires Adjustment, Insertion Tube - Dull and Worn, Light Transmission - 1 Light Guide Glass Damaged, Channels - No Fault Found, Leak Check - Leak Present *2400242* View details Winning BidN/A Lot 11 Olympus GIF-Q260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Stiff, Requires Adjustment, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Slow Leak Present *2756488* View details Winning BidN/A Lot 12 Olympus BF-260 Video Bronchoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Stiff, Down Not Reaching Specification, Up Requires Adjustment, Insertion Tube - Buckled at Control Body, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Slow Leak Present *2902344* View details StatusUnsold Lot 14 Olympus CF-H260DL Video Colonoscope in Case - Engineer's Report : Optical System - Slightly Dark in Center, Angulation - Up / Down Loose, Requires Adjustment, Insertion Tube - No Fault Found, Light Transmission - 1 Light Bundle 60% Loss, Channels - No Fault Found, Leak Check - No Fault Found *2800262* View details Winning BidN/A Lot 15 Olympus CF-H260DL Video Colonoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2200623* View details Winning BidN/A Lot 16 Olympus CF-H260DL Video Colonoscope in Case - Engineer's Report : Optical System - Cloudy in Top Left, Angulation - Some Snaking in Bending Section, Tight, Requires Adjustment, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2800242* View details Winning BidN/A Lot 17 Olympus GIF-Q260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2745244* View details Winning BidN/A Lot 18 Olympus CF-Q260DL Video Colonoscope in Case - Engineer's Report : Optical System - No Image, Angulation - Tight, Requires Adjustment, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2710638* View details Winning BidN/A Lot 19 Olympus CF-Q260DL Video Colonoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - Slight Light Loss, Channels - Clear, Leak Check - No Fault Found *2610403* View details Winning BidN/A Lot 20 Olympus GIF-Q240Z Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2200175* View details Winning BidN/A Lot 21 Olympus BF-260 Video Bronchoscope in Case - Engineer's Report : Optical System - Small Cloudy Spot in Centre, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - Clear, Leak Check - No Fault Found *2902341* View details Winning BidN/A Lot 22 Olympus CF-H260DL Video Colonoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found, Other Comments - Light Buckle at Control Boot *2400904* View details Winning BidN/A Lot 23 Olympus CF-H260DL Video Colonoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found, Other Comments - UPD Protective Cap Missing *2800234* View details Winning BidN/A Lot 24 Olympus CF-Q260DL Video Colonoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found, Other Comments - UPD Cap Missing *2610381* View details Winning BidN/A Lot 26 Olympus CF-H260DL Video Colonoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Like New, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2200625* View details Winning BidN/A Lot 27 Olympus GIF-XQ260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2421551* View details Winning BidN/A Lot 30 Olympus ENF-P4 Pharyngoscope - Engineer's Report : Optical System - 0 Broken Fibers, Cloudy View, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - No Fault Found *W911010* (H) View details Winning BidN/A Lot 31 Olympus ENF-P4 Pharyngoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - No Fault Found *W1429832* (H) View details Winning BidN/A Lot 32 Olympus ENF-P4 Pharyngoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - No Fault Found *W911020* (H) View details Winning BidN/A Lot 34 Olympus BF Type P20D Bronchoscope - Engineer's Report : Optical System - No Fault Found, Angulation - Down out of Spec, Insertion Tube - Buckled at Control Boot, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2002132* View details Winning BidN/A Lot 35 Karl Storz 11272C1 Cystoscope in Case - Engineer's Report : Optical System - 10 Broken Fibers, Angulation - No Fault Found, Insertion Tube - Light Buckle at Control Body, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Untested, Other Comments - Crush at Bending Section *2016938S* View details Winning BidN/A Lot 36 Olympus ENF-P4 Pharyngoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - No Fault Found *W911019* (H) View details Winning BidN/A Lot 37 Olympus ENF-P4 Pharyngoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - No Fault Found *1911498* (H) View details Winning BidN/A Lot 38 Olympus ENF-P4 Pharyngoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - No Fault Found *W911011* (H) View details Winning BidN/A Lot 39 Pentax EB-1570K Video Bronchoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *G121980* View details Winning BidN/A Lot 40 Pentax EB-1970K Video Bronchoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *G121475* View details Winning BidN/A Lot 41 Pentax EC-3890Fi Video Colonoscope in Case - Engineer's Report : Optical System - Unable to Check Image, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *A110915* View details Winning BidN/A Lot 42 Vision Sciences TNE-2000 Rhinolaryngoscope in Case - Engineer's Report : Optical System - Multiple Broken Fibers, Angulation - Snaking, Insertion Tube - Multiple Crushed Sections, Light Transmission - Loss of Light, Channels - Blocked, Leak Check - No Fault Found, Other Comments - Light Guide Coating Damaged *T02745* (H) View details StatusUnsold Lot 43 Vision Sciences TNE-2000 Rhinolaryngoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Requires Adjustment, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - Leak Present from Bending Section Rubber, Other Comments - Crushed Light Tube *T0278S* (H) View details StatusUnsold Lot 44 Vision Sciences TNE-5000 Rhinolaryngoscope in Case - Engineer's Report : Optical System - Untested, Angulation - Damaged Bending Section, Insertion Tube - Bending Section Damaged, No Bending Section Rubber, Light Transmission - No Fault Found, Leak Check - Untested But Suspect Leak will be Present *TE30116P* (H) View details StatusUnsold Lot 45 Vision Sciences TNE-5000 Rhinolaryngoscope in Case - Engineer's Report : Optical System - Untested, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - Untested *TE40113P* (H) View details StatusUnsold Lot 46 Vision Sciences TNE-5000 Rhinolaryngoscope in Case - Engineer's Report : Optical System - Untested, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - No Fault Found, Other Comments - Damage to Parts on Control Body *350108* (H) View details StatusUnsold Lot 47 Olympus BF Type 10 Bronchoscope in Case - Engineer's Report : Optical System - 33 Broken Fibers, Angulation - No Fault Found, Insertion Tube - Badly Kinked, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2511726* View details Winning BidN/A Lot 48 Olympus CF-Q240AL Video Colonoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2010060* View details Winning BidN/A Lot 49 Olympus GIF-Q240 Video Gastroscope - Engineer's Report : Optical System - No Fault Found, Angulation - Left / Right Brake Faulty, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2030852* View details Winning BidN/A Lot 50 Olympus GIF-XQ240 Video Gastroscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2930059* View details Winning BidN/A Lot 51 Olympus CF-240AL Video Colonoscope - Engineer's Report : Optical System - Black Mark on Image, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2010137* View details StatusUnsold Lot 52 Olympus GIF-Q260 Video Gastroscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2644952* View details Winning BidN/A Lot 53 Olympus CF-240L Video Colonoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2900338* View details Winning BidN/A Lot 54 Olympus GIF-XQ240 Video Gastroscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2033711* View details Winning BidN/A Lot 55 Olympus CF-240AL Video Colonoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Intermittent Leak Present *2110207* View details StatusUnsold Lot 56 Olympus GIF-XQ240 Video Gastroscope - Engineer's Report : Optical System - No Fault Found, Angulation - Tension in Motion, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2133950* View details Winning BidN/A Lot 57 Olympus CF-Q260DL Video Colonoscope - Engineer's Report : Optical System - Parallel Lines on Image, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2710666* View details Winning BidN/A Lot 58 Olympus TJF-240 Video Duodenoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2800196* View details Winning BidN/A Lot 59 Olympus BF Type 200 Video Bronchoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2500906* View details Winning BidN/A Lot 60 Olympus GIF-Q260 Video Gastroscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Minor Indents Present, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2745283* View details Winning BidN/A Lot 61 Pentax ECY-1570K Video Cystoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Crush and Kink Marks Present, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *G120330* View details Winning BidN/A Lot 62 Pentax EG-2990i Video Gastroscope in Case - Engineer's Report : Optical System - Untested, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *A115906* View details Winning BidN/A Lot 63 Pentax EG-2990i Video Gastroscope in Case - Engineer's Report : Optical System - Untested, Angulation - Strained, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *A115610* View details Winning BidN/A Lot 64 Pentax FNL-10RBS Laryngoscope with BS-LC1 Light Cable - Engineer's Report : Optical System - 3 Broken Fibers and Misty Optics, Angulation - No Fault Found, Insertion Tube - Crush and Kink Marks Present, Light Transmission - No Fault Found, Other Comments - Light Guide Tube Badly Torn *G111483* View details Winning BidN/A Lot 65 DP Medical TF6 Pharyngoscope with Leak Tester and Light Cable in Case - Engineer's Report : Optical System - Faint Stain, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - No Fault Found (H) View details Winning BidN/A Lot 66 Olympus ENF-P4 Pharyngoscope in Case - Engineer's Report : Optical System - 1 Broken Fiber, Angulation - No Fault Found, Insertion Tube - Kinked at Grip, Light Transmission - No Fault Found, Channels - No Fault Found *1125399* (H) View details Winning BidN/A Lot 67 Olympus ENF-P4 Pharyngoscope in Case - Engineer's Report - Optical System - 0 Broken Fibers, Leak at Eyepiece, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - Leak at Eyepiece *1911398* (H) View details Winning BidN/A Lot 69 Olympus ENF-P4 Pharyngoscope - Engineer's Report : Optical System - 4 Broken Fibers, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - No Fault Found *W911014* (H) View details StatusUnsold Lot 72 Pentax FI-10BS Laryngoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Leak within Bending Section *G111111* View details Winning BidN/A Lot 73 Pentax EB-1570K Video Bronchoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *A120185* View details Winning BidN/A Lot 74 Pentax EB-1970K Video Bronchoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Play in Controls and Strained in Use, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Internal Leak Present *A120176* View details Winning BidN/A Lot 75 Pentax EB-1570K Video Bronchoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *G121954* View details Winning BidN/A Lot 76 Pentax FI-10BS Laryngoscope in Case - Engineer's Report : Optical System - Fluid in Optics, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *A110539* View details Winning BidN/A Lot 77 Olympus GIF-XQ230 Video Gastroscope in Case - Engineer's Report : Optical System - No Image, Angulation - All Movements Inoperative, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2603049* View details Winning BidN/A Lot 78 Pentax FI-10BS Laryngoscope in Case - Engineer's Report : Optical System - 4 Broken Fibers, Angulation - No Fault Found, Insertion Tube - Crush Mark Present, Light Transmission - Poor, Leak Check - Excessive Leak Present *G111150* View details Winning BidN/A Lot 79 Vision Sciences TNE-2000 Rhinolaryngoscope in Case - Engineer's Report : Optical System - 2 Broken Fibers, Angulation - No Fault Found, Insertion Tube - Badly Crushed, Light Transmission - No Fault Found, Leak Check - Light Guide Tube Torn and Leaking *T0226P* (H) View details StatusUnsold Lot 80 Vision Sciences TNE-5000 Rhinolaryngoscope in Case - Engineer's Report : Optical System - Untested, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - Untested *TE40117P* (H) View details StatusUnsold Lot 81 Olympus CF-H260DL Video Colonoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2800079* View details Winning BidN/A Lot 82 Olympus CF-240DL Video Colonoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2400217* View details Winning BidN/A Lot 83 Olympus CF-H260DL Video Colonoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2800078* View details Winning BidN/A Lot 85 Olympus GIF-Q260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2150398* View details Winning BidN/A Lot 86 Olympus CF-240DL Video Colonoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Minor Ripple, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2200050* View details Winning BidN/A Lot 87 Olympus PCF-Q260AL Video Colonoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2610213* View details Winning BidN/A Lot 91 Olympus CF-240DL Video Colonoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Left / Right Brake Faulty, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2500227* View details Winning BidN/A Lot 92 Olympus GIF-Q260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2421692* View details Winning BidN/A Lot 94 Olympus CF-240DL Video Colonoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2300109* View details Winning BidN/A Lot 95 Olympus GIF-Q260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2150401* View details Winning BidN/A Lot 96 Olympus ENF-VQ Video Rhinolaryngoscope in Case - Engineer's Report : Optical System - Untested, Angulation - No Fault Found, Insertion Tube - Crush / Kink Mark Present, Light Transmission - No Fault Found, Leak Check - Excessive Body Leak *W040104* View details Winning BidN/A Lot 97 Olympus CF-240DL Video Colonoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2300108* View details StatusUnsold Lot 99 Olympus GIF-Q260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2421748* View details Winning BidN/A Lot 100 Karl Storz 11272VP Video Urethro-Cystoscope in Case - Engineer's Report : Optical System - Untested, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *21809* View details Winning BidN/A Lot 101 Karl Storz 11272VP Video Urethro-Cystoscope with Leak Tester in Case - Engineer's Report : Optical System - Untested, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *15657* View details Winning BidN/A Lot 102 Karl Storz 11272VP Video Urethro-Cystoscope in Case - Engineer's Report : Optical System - Untested, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *17484* View details Winning BidN/A Lot 103 Karl Storz 11272VP Video Urethro-Cystoscope with Leak Tester and Accessories in Case - Engineer's Report : Optical System - Untested, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *22192* View details Winning BidN/A Lot 104 Karl Storz 11272VP Video Urethro-Cystoscope in Case - Engineer's Report : Optical System - Untested, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *66794* View details Winning BidN/A Lot 105 Karl Storz 11272VP Video Urethro-Cystoscope with Leak Tester and Accessories in Case - Engineer's Report : Optical System - Untested, Angulation - Strained, Insertion Tube - Bending Section Severely Crushed, Light Transmission - 20% Loss, Channels - Restricted, Leak Check - Leak from Tube *11084* View details Winning BidN/A Lot 106 Pentax EG-3870UTK Ultrasound Video Gastroscope in Case (Spares and Repairs) *A120479* View details Winning BidN/A Lot 107 Olympus ENF-P4 Pharyngoscope - Engineer's Report : Optical System - 5 Broken Fibers and Slight Stain Present, Angulation - No Fault Found, Insertion Tube - Kinked and Crushed, Light Transmission - No Fault Found, Leak Check - No Fault Found *1901215* View details Winning BidN/A Lot 108 Olympus ENF-P4 Pharyngoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Minor Kink Present, Light Transmission - No Fault Found, Leak Check - No Fault Found *2723821* View details Winning BidN/A Lot 109 Olympus ENF-P4 Pharyngoscope - Engineer's Report : Optical System - 5 Broken Fibers, Angulation - No Fault Found, Insertion Tube - Kinked / Wrinkled, Light Transmission - No Fault Found, Leak Check - No Fault Found *2723833* View details StatusUnsold Lot 110 Olympus ENF-P4 Pharyngoscope - Engineer's Report : Optical System - 3 Broken Fibers and Fluid Stain Present, Angulation - No Fault Found, Insertion Tube - Wrinkled, Leak Check - No Fault Found *W700231* View details StatusUnsold Lot 111 Olympus ENF-P4 Pharyngoscope - Engineer's Report : Optical System - 1 Broken FIber, Angulation - No Fault Found, Insertion Tube - Wrinkled and Crush Mark Present, Light Transmission - No Fault Found, Leak Check - No Fault Found *2723836* View details StatusUnsold Lot 112 Olympus LF-GP Laryngoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *W301409* View details Winning BidN/A Lot 113 Optimed 800.225.7486 Optim Entity NasoView LED Incorporated Pharyngoscope with 3 x Batteries, 1 x Battery Charger and Leak Tester in Case (Powers Up) - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found View details Winning BidN/A Lot 114 Pentax FI-13P Laryngoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *A01002* View details StatusUnsold Items per page 102550100 Previous|12|Next Previous 12 Next Previous 12 Next