BRITISH MEDICAL AUCTIONS December 2021 Endoscopy (#461) 09/12/2021 2:00 PM GMT - 17/12/2021 2:08 PM GMT CLOSED Starts Ending 17/12/2021 1:00 PM GMT Auction info BRITISH MEDICAL AUCTIONS View Search ▼ Items per page 102550100 Search Sort by DefaultTime LeftLot#Lot NameNewestHighest PriceLowest PriceBidsViews Filter By Categories Category match ANY of the aboveALL of the above Featured Exclude closed lots Price Lot Number Results: Viewing items 1-33 of 33. Previous|1|Next Lot 1 Reichert Fiber Optics VFS2510009 Endoscope - Engineer's Report : Optical System - 3 Broken Fibers and Minor Dirt, Angulation - No Fault Found, Insertion Tube - Kinked and Strained, Light Transmission - No Fault Found, Channels - Unable to Check, Leak Check - Unable to Check View details Winning BidN/A Lot 2 Olympus CF-T10L Colonoscope - Engineer's Report : Optical System - Excessive Broken Fibers and Severe Fluid Stain, Angulation - Not Reaching Specification, Play in Controls, Bending Section Strained, Insertion Tube - Severely Kinked, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Leak Present View details Winning BidN/A Lot 3 Olympus ENF-P4 Pharyngoscope - Engineer's Report : Optical System - 7 Broken Fibers, Angulation - No Fault Found, Insertion Tube - Crush and Kink Marks Present, Light Transmission - No Fault Found, Leak Check - No Fault Found *W800549* View details Winning BidN/A Lot 4 Olympus ENF-P4 Pharyngoscope - Engineer's Report : Optical System - 3 Broken Fibers, Angulation - No Fault Found, Insertion Tube - Kinked, Light Transmission - No Fault Found, Leak Check - No Fault Found *W800570* View details Winning BidN/A Lot 5 Olympus ENF-P4 Pharyngoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Kinked, Light Transmission - No Fault Found, Leak Check - No Fault Found *W800579* View details Winning BidN/A Lot 6 Olympus ENF-P4 Pharyngoscope - Engineer's Report : Optical System - 4 Broken Fibers, Angulation - No Fault Found, Insertion Tube - Crush and Kink Marks Present, Light Transmission - No Fault Found, Leak Check - No Fault Found *W800085* View details Winning BidN/A Lot 7 Olympus ENF-P4 Pharyngoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Crush Mark Present, Light Transmission - No Fault Found, Leak Check - No Fault Found *W211943* View details Winning BidN/A Lot 8 Olympus GIF-XQ20 Gastroscope - Engineer's Report : Optical System - 14 Broken Fibers and Stain Present, Angulation - No Fault Found, Insertion Tube - No Fault Found, Minor Bite Marks, Light Transmission - Stained, Channels - No Fault Found, Leak Check - Excessive Leak Present *2119553* View details Winning BidN/A Lot 9 Olympus GIF-XQ230 Video Gastroscope - Engineer's Report : Optical System - Untested Due to Damaged Hub, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *S/N Unreadable* View details Winning BidN/A Lot 10 Olympus CHF-10 Choledocoscope - Engineer's Report : Optical System - 10 Broken Fibers, Angulation - No Fault Found, Insertion Tube - Kinked, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *1510155* View details Winning BidN/A Lot 11 Olympus CYF-4A Cystoscope - Engineer's Report : Optical System - Excessive Fluid Contamination, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *W000617* View details Winning BidN/A Lot 12 Olympus CYF-5A Cystoscope - Engineer's Report : Optical System - Excessive Dirt Particles, Angulation - Strained, Lever Broken, Not Reaching Specification, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *W800730* View details Winning BidN/A Lot 13 Olympus CYF-5A Cystoscope - Engineer's Report : Optical System - 15 Broken Fibers and Minor Stain, Angulation - No Fault Found, Insertion Tube - Badly Kinked, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *W600374* View details Winning BidN/A Lot 14 Olympus CYF-5A Cystoscope - Engineer's Report : Optical System - 1 Broken Fiber, Angulation - No Fault Found, Insertion Tube - Minor Crush Mark, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *W800739* View details Winning BidN/A Lot 15 Olympus ENF-GP Pharyngoscope in Case - Engineer's Report : Optical System - 16 Broken Fibers, Angulation - Crush in Bending Section, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - Bending Section Rubber Leaking *2713288* (G) View details Winning BidN/A Lot 16 Olympus ENF-GP Pharyngoscope with Olympus WA91502A Mini Light Source (Untested Due to No Batteries) in Case - Engineer's Report : Optical System - 6 Broken Fibers, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - No Fault Found *2713268* (G) View details Winning BidN/A Lot 17 Olympus ENF-GP Pharyngoscope in Case - Engineer's Report : Optical System - 3 Broken Fibers, Angulation - Bending Section Crushed, Bending Section Rubber Perforated, Insertion Tube - Crush Mark, Light Transmission - No Fault Found, Leak Check - Bending Section Perforated *W901769* (G) View details Winning BidN/A Lot 18 Olympus ENF-P4 Pharyngoscope in Case - Engineer's Report - Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Minor Crush Marks, Light Transmission - No Fault Found, Leak Check - No Fault Found *W500005* (G) View details Winning BidN/A Lot 19 Olympus ENF-GP Pharyngoscope in Case - Engineer's Report : Optical System - 7 Broken Fibers and Fluid Stain, Angulation - No Fault Found, Insertion Tube - Peeling, Light Transmission - No Fault Found, Leak Check - Excessive Leak Present *W801093* (G) View details Winning BidN/A Lot 20 Olympus TJF-260V Video Duodenoscope in Case - Engineer's Report : Optical System - Mark on Image, Angulation - Play Not to Spec, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - Raiser Wire Fraying, Leak Check - No Fault Found *2001296* View details Winning BidN/A Lot 21 Pentax ES-3870K Sigmoidoscope in Case - Engineer's Report : Optical System - Unable to Check, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - Unable to Check, Leak Check - Unable to Check *A120031* View details Winning BidN/A Lot 22 Olympus GIF-XQ260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Bending Section Strained, Insertion Tube - Minor Kink, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2300723* View details Winning BidN/A Lot 23 Pentax FNL-10RBS Laryngoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Crush Mark and Kink Present, Light Transmission - No Fault Found *4115976* View details Winning BidN/A Lot 24 Pentax FNL-10RBS Laryngoscope in Case - Engineer's Report : Optical System - 15 Broken Fibers and Stain Present, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - Leaking, Other Comments - Partly Dismantled and Parts Missing *4115636* View details Winning BidN/A Lot 25 Pentax FNL-10RBS Laryngoscope in Case - Engineer's Report : Optical System - Excessive Broken Fibers, Angulation - Broken, Insertion Tube - Crush Marks, Light Transmission - No Fault Found, Other Comments - Partly Dismantled *415965* View details Winning BidN/A Lot 26 Pentax FNL-10RBS Laryngoscope in Case - Engineer's Report : Optical System - Excessive Broken Fibers, Angulation - No Fault Found, Insertion Tube - Minor Crush Marks, Light Transmission - No Fault Found, Leak Check - Leaking, Other Comments - Body Crack, Scope Partly Dismantled *A110632* View details Winning BidN/A Lot 27 Olympus BF-260 Video Bronchoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Badly Crushed, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2701451* View details Winning BidN/A Lot 28 Karl Storz 13801PKS Video Gastroscope in Case - Engineer's Report : Optical System - Unable to Check, Angulation - Not Reaching Specification, To Be Adjusted, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - Unable to Check, Leak Check - Unable to Check View details Winning BidN/A Lot 29 Olympus LF-V Video Laryngoscope in Case - Engineer's Report : Optical System - Image Fault, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - Unable to Check, Leak Check - Excessive Leak Present *2010324* View details Winning BidN/A Lot 30 Olympus CF-Q260DL Video Colonoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Play in Controls, Not Reaching Specification, To Be Adjusted, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2911208* View details Winning BidN/A Lot 31 Olympus GIF-XQ260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Not Reaching Specification, To Be Adjusted, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2310781* View details Winning BidN/A Lot 32 Olympus TJF-260V Video Duodenoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Rubber Grip Torn, Light Transmission - No Fault Found, Channels - Raiser Bridge Op Broken, Leak Check - No Fault Found *2700313* View details Winning BidN/A Lot 33 Olympus CF-Q260DL Video Colonoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Not Reaching Specification, To Be Adjusted, Insertion Tube - Bending Section Rubber Glue Poor, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2710859* View details Winning BidN/A Items per page 102550100 Previous|1|Next Previous 1 Next Previous 1 Next