BRITISH MEDICAL AUCTIONS November 2021 Endoscopy Equipment (#440) 26/10/2021 5:00 PM BST - 02/11/2021 2:26 PM GMT CLOSED Starts Ending 02/11/2021 1:02 PM GMT View details Auction info BRITISH MEDICAL AUCTIONS Search Sort by DefaultTime LeftLot#Lot NameNewestHighest PriceLowest PriceBidsViewsDefaultDefaultTime LeftLot#Lot NameNewestHighest PriceLowest PriceBidsViews Lot Number Filter By Categories Category match ANY of the aboveALL of the aboveANY of the aboveANY of the aboveALL of the above Featured Price Exclude closed lots View Search ▼ Results: Viewing items 1-50 of 50. Previous|1|Next Items per page 102550100100 Lot 1 Olympus CYF-5 Cystoscope in Case - Engineer's Report : Optical System - 1 Broken Fiber, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *W147629* View details Winning BidN/A Lot 2 Olympus ENF-P4 Pharyngoscope in Case - Engineer's Report : Optical System - 1 Broken Fiber, Angulation - Minor Kink at Grip, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Leak Check - No Fault Found *1227231* View details Winning BidN/A Lot 3 Olympus ENF-P4 Pharyngoscope in Case - Engineer's Report : Optical System - 4 Broken Fibers, Angulation - No Fault Found, Insertion Tube - Kink at Grip, Light Transmission - No Fault Found, Leak Check - No Fault Found *W312054* View details Winning BidN/A Lot 4 Fujinon EB-530US Ultrasound Video Bronchoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Kink at Grip, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Leak Present, Other Comments - Unable to Check Ultrasound Quality *RU912A037* View details Winning BidN/A Lot 5 Fujinon EB-530US Ultrasound Video Bronchoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Not Reaching Specification, To Be Adjusted, Insertion Tube - Kinked at Grip, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Excessive Leak Present, Other Comments - Unable to Check Ultrasound Quality *921A003* View details Winning BidN/A Lot 6 Olympus GIF-FQ260Z Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2910320* View details Winning BidN/A Lot 7 Olympus CYF-5 Cystoscope in Case - Engineer's Report : Optical System - 0 Broken Fibers, Minor Dirt on Mask, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *W704323* View details Winning BidN/A Lot 8 Olympus CYF-5 Cystoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *W147860* View details Winning BidN/A Lot 9 Olympus CYF-5 Cystoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *W805795* View details Winning BidN/A Lot 10 Olympus GIF-Q260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Not Reaching Specification, To Be Adjusted, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2856686* View details Winning BidN/A Lot 11 Olympus GIF-Q260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Not Reaching Specification, To Be Adjusted, Insertion Tube - Kinks at Grip, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2745436* View details Winning BidN/A Lot 12 Olympus CYF-5 Sterrad Edition Cystoscope in Case - Engineer's Report : Optical System - 1 Broken Fiber, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *W247983* View details Winning BidN/A Lot 13 Olympus CYF-5 Cystoscope in Case - Engineer's Report : Optical System - 1 Broken Fiber, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *147862* View details Winning BidN/A Lot 14 Olympus GIF-Q260 Video Gastroscope in Case -Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - Excessive Bio Leak Present, Leak Check - Excessive Bio Leak Present *2745438* View details Winning BidN/A Lot 15 Olympus ENF-L3 Pharyngoscope in Case - Engineer's Report : No Fault Found, Angulation - No Fault Found, Insertion Tube - Split at Prox, Light Transmission - No Fault Found, Leak Check - Leak Insertion Tube *1410202* View details StatusUnsold Lot 16 Olympus GF-UCT240 Ultrasound Video Gastroscope in Case - Engineer's Report : Optical System - Unable to Check, Angulation - Not Reaching Specification, To Be Adjusted, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found *1600071* View details Winning BidN/A Lot 17 Olympus GF-UCT240 Ultrasound Video Gastroscope in Case - Engineer's Report : Optical System - Unable to Check, Angulation - L-R Brake Jump, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found *1600071* View details Winning BidN/A Lot 18 Olympus CYF-5 Cystoscope in Case - Engineer's Report : Optical System - 0 Broken Fibers, Minor Dirt on Mask, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *W247985* View details StatusUnsold Lot 19 Olympus MH-908 Ultrasound Probe Endoscope in Case - Engineer's Report : Optical System - Unable to Check, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - Unable to Check, Leak Check - Unable to Check *1110148* View details StatusUnsold Lot 20 Olympus GIF-Q260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - Not Reaching Specification, To Be Adjusted, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2745420* View details Winning BidN/A Lot 21 Olympus OVC-200 Lecture Scope Attachment in Case - Engineer's Report : Optical System - No Fault Found, Insertion Tube - No Fault Found View details Winning BidN/A Lot 22 Pentax FI-10P2 Laryngoscope - Engineer's Report : Optical System - No Eyepiece, Angulation - No Fault Found, Insertion Tube - Strained, Light Transmission - No Fault Found, Channels - Restricted, Leak Check - Excessive Leak, Other Comments - Scope Partially Dismantled, Spares and Repairs *A01656* View details Winning BidN/A Lot 23 Pentax FNL-10RBS Pharyngoscope - Engineer's Report : Optical System - 2 Broken Fibers, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *G112616* View details Winning BidN/A Lot 24 Olympus GIF-H260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2734599* View details Winning BidN/A Lot 25 Olympus CYF-5 with Light Guide Tube in Case - Engineer's Report : Optical System - 2 x Broken Fibers, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *W247921* View details Winning BidN/A Lot 26 Olympus CYF-5 with Light Guide Tube in Case - Engineer's Report : Optical System - 8 x Broken Fibers, Angulation - Short of Specification, Insertion Tube - Poly Coating Scuffed, Light Transmission - No Fault Found, Channels - Catch in Channel - Leak Check - No Fault Found *W247883* View details Winning BidN/A Lot 27 Olympus GIF-Q260 Video Gastroscope in Case -Engineer's Report : Optical System - No Fault Found, Angulation - Not Reaching Specification, To Be Adjusted, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - Excessive Bio Leak Present, Leak Check - Excessive Bio Leak Present *2856768* View details Winning BidN/A Lot 28 Olympus CYF-5 Sterrad Ready Cystoscope in Case - Engineer's Report : Optical System - 1 Broken Fiber, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *W247925* View details Winning BidN/A Lot 29 Olympus GIF-1T240 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2230520* View details Winning BidN/A Lot 30 Olympus GIF-H260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2734558* View details Winning BidN/A Lot 31 Olympus GIF-H260 Video Gastroscope in Case - Engineer's Report : Optical System - Minor Parallel Line in Interface Across Image, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2734556* View details Winning BidN/A Lot 32 Olympus GIF-Q260 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Crush Mark at 40cm, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2644163* View details Winning BidN/A Lot 33 Olympus GIF-1T240 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2830426* View details Winning BidN/A Lot 34 Olympus CYF-5 with Light Guide Tube in Case - Engineer's Report : Optical System - 2 x Broken Fibers, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *W805340* View details Winning BidN/A Lot 35 Olympus LF-V Video Laryngoscope - Engineer's Report : Optical System - Unable to Check, Angulation - Up Inoperative, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Unable to Check *2710180* View details Winning BidN/A Lot 36 Pentax ES-3830K Video Sigmoidoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *E12093* View details Winning BidN/A Lot 37 Olympus BF-260 Video Bronchoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Kink Present, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2501086* View details Winning BidN/A Lot 38 Olympus BF-UC260-0LE Video Ultrasound Bronchoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found, Other Comments - Unable to Check Ultrasound Imaging *1600071* View details Winning BidN/A Lot 39 Olympus LF-GP Laryngoscope - Engineer's Report : Optical System - No Fault Found, Angulation - Not Reaching Specification, Insertion Tube - Minor Indentations, Light Transmission - No Fault Found, Channels - No Fault Found *W101172* View details Winning BidN/A Lot 40 Olympus CF-Q260DL Video Colonoscope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Bending Rubber and Mesh Missing, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Unable to Check *2311985* View details Winning BidN/A Lot 41 Pentax FCP-9P 2 Meter Scope - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Bad Crush Marks and Tube Sleeving Damage, Light Transmission - No Fault Found, Channels - Catch, Leak Check - Unable to Check View details Winning BidN/A Lot 100 Olympus CYF-5 Cystoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *W906494* View details Winning BidN/A Lot 101 Olympus CYF-5 Cystoscope in Case - Engineer's Report : Optical System - 1 Broken Fiber, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *W247890* View details Winning BidN/A Lot 102 Olympus CYF-5 Cystoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *W601813* View details Winning BidN/A Lot 103 Olympus GIF-2T240 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2231118* View details Winning BidN/A Lot 104 Olympus GIF-2T240 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2231150* View details Winning BidN/A Lot 105 Olympus GIF-2T240 Video Gastroscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2530793* View details Winning BidN/A Lot 106 Olympus CF-FH260AZL Video Colonoscope in Case - Engineer's Report - Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - No Fault Found, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - No Fault Found *2700106* View details Winning BidN/A Lot 107 Fujinon EB-530US Video Ultrasound Bronchoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Leaking Tube, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Leak Present *912A036* View details Winning BidN/A Lot 108 Fujinon EB-530US Video Ultrasound Bronchoscope in Case - Engineer's Report : Optical System - No Fault Found, Angulation - No Fault Found, Insertion Tube - Leaking Tube, Light Transmission - No Fault Found, Channels - No Fault Found, Leak Check - Leak Present *912A004* View details Winning BidN/A Items per page 102550100100 Previous|1|Next Previous 11 Next Previous 11 Next